Illinois Reuse Machineries and Industrial Equipment Newsboad   >   Michigan   >   Warren   >   Valleylab electrosurgical generator force 2-20PC

Valleylab electrosurgical generator force 2-20PC

VALLEYLAB ELECTROSURGICAL GENERATOR FORCE 2-20PC
UNIT IS PATIENT READY- NICE AND CLEAN, SEE PICTURES AND CALIBRATION RESULTS FOR YOUR GUARANTEED SATISFACTION WITH 14 DAY REFUND PLUS 45 DAY WARRANTY.
The Valleylab Force 2 ESU electrosurgical cautery unit provides cutting and coagulation in monopolar or bipolar outputs. Force 2s have audible and visual alarms, including Fault Indicator. The REM (Return Electrode Monitoring) is standard. 300 watts max. output in a Cut setting, 70 watts max. output in Bipolar setting, 120 watts max. output in Coagulation setting and 250 Watts max. output in a Blend 1 setting, 200 Watts max. output in Blend 2 setting, and 150 Watts max. in Blend 3 setting.
- 1 Reusable Patient Plate with Valleylab REM Adaptor & Cable
CALIBRATION TEST RESULTS : BEFORE ADJUSTMENT : Manufacturer's specification : +/- 15 %
ELECTROSURGICAL GENERATOR : FORCE 2-20PC
Monopolar @ 300 Ohms rated load. Bipolar @ 100 Ohms rated load.
Setting : Pure Cut / Blend 1 / Blend 2 / Blend 3 Setting : Coag Output Setting : Bipolar Output
300 280.3 - - - 120 121.4 70 70.9
250 241.5 239.2 - - 100 105.5 60 58.7
200 195.2 193.6 191.1 - 80 90.3 50 51.0
150 149.8 150.5 146.4 141.5 60 69.0 40 40.7
100 105.3 106.1 102.2 99.0 40 45.3 30 31.2
60 62.9 63.2 63.7 60.7 20 23.0 20 20.1
40 43.9 44.9 45.3 43.8 10 11.0 10 9.6
Leakage current (Chassis) 28 microamp. 24 microamp.
Leakage current (Earth) 28 microamp. 24 microamp.
Chassis Resistance Measured : 0.22 Ohm
Valleylab electrosurgical generator force 2-20PC Valleylab electrosurgical generator force 2-20PC Valleylab electrosurgical generator force 2-20PC Valleylab electrosurgical generator force 2-20PC Valleylab electrosurgical generator force 2-20PC Valleylab electrosurgical generator force 2-20PC Valleylab electrosurgical generator force 2-20PC Valleylab electrosurgical generator force 2-20PC